Journal of Mechanical Engineering ›› 2016, Vol. 52 ›› Issue (8): 58-66.doi: 10.3901/JME.2016.08.058

Previous Articles     Next Articles

Fault-tolerant Measurement Mechanism and Calibration Experimental Study of Pre-stressed Parallel Six-axis Force Sensor

YAO Jiantao1, 2, CUI Pengxiao1, ZHU Jialong1, XU Yundou1, 2, ZHAO Yongsheng1, 2   

  1. 1.Parallel Robot and Mechatronic System Laboratory of Hebei Province, Yanshan University, Qinhuangdao 066004;
    2. Key Laboratory of Advanced Forging & Stamping Technology and Science, Ministry of Education of China, Yanshan University, Qinhuangdao 066004
  • Online:2016-04-15 Published:2016-04-15

Abstract: For a pre-stressed six-axis force sensor, the theoretical analysis and experimental study of fault-tolerant measurement mechanism are carried out. The fault-tolerance characteristic of pre-stressed parallel six-axis force sensor is introduced, with the screw theory and deformation coordination, the mathematic models including fault-free mathematic model and signal fault mathematic model are established, respectively. Moreover, the fault-tolerant measurement principle is revealed. The static calibration algorithm of the pre-stressed parallel six axis force sensor is analyzed, a fault-tolerant calibration method is proposed and the static calibration includes unidirection loading, bidirectional combined loading and sectional loading of the sensor is carried out including the models with and without signal fault. Through analyzing the experimental data, the linear error of the sensor is drawn under different working conditions. The research lays a theoretical and experimental foundation for the application of the fault-tolerant six-axis force sensor.

Key words: calibration experiment, fault-tolerant, parallel, pre-stressed, six-axis force sensor